CLIA-Validated Whole Blood Mass Cytometry Immune Profiling Assay

Clinical trials-grade immunophenotyping with 40+ markers for whole blood immune detection

Identify immune biomarkers for dose selection, mechanism of action, and pharmacodynamic studies in your clinical trials.

Analyze up to 735 immune cell subsets and states with our 41-marker panel. Accurate subset measurements with <10% coefficient of variation for all major immune cell types.

Total Average Coefficient of Variation:

  • Intra-Run Precision: 2.18%
  • Inter-Run Precision: 6.84%
  • Inter-Kit Precision: 5.75%

Validated with our TokuKit – designed for simplified sample collection at clinical sites.

Up to 720+ Immune Subsets and States Analyzed

Page Anchor: #whole-blood-mass-cytometry – Not Visible On Front Page Anchor: #whole-blood-mass-cytometry-validation – Not Visible On Front